616 Hollywood Avenue, Cherry Hill, NJ 08002
(856) 663-4700 Toll Free (800) 333-3131 |
||||||||
| by Mizzy A Division of Keystone Industries |
||||||||
| Lower Anterior Infiltration Anesthesia
*Always administer an initial .05cc injection to provide subtopical anesthesia for painless application. Place the SyriJet just below the ridge, at the mucal gingival junction. Procedure: Cavity Preparation
Lift lip and dry the tissue. Place the tip of the SyriJet as high as possible on the attached tissue, GENTLY in contact and at RIGHT ANGLES to the surface of the tissue. Do Not compress the tissue; Do Not exert pressure. Drop the lip. Hold the instrument motionless and gently squeeze the trigger. Do not lift the lip immediately after injection. Extractions: Permanent Lower Anteriors and All Primary Teeth These procedures require two (2) injections. First injection - Apex infiltration (same as above)
Second injection is a lingual injection with Rubber Cap Removed. (see third picture on left) SyriJet tip is placed approximately 1-3 millimeters above the gingival margin.
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
![]() |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Next Page |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||